Technical Information

Further information regarding any of the topics presented here is provided
in Installation Planning and System Maintenance Guide (A-61000).

Installation

This equipment and accessories are installed by a Kodak Field Engineer
and takes approximately 4 hours.

Acoustical emissions

Scanner 990 — Acoustical emissions shall be less than or equal to 67 dB
during operation and less than or equal to 62 dB while idle.

Scanner 9500 — Acoustical emissions shall be less than or equal to 76 dB
during operation and less than or equal to 67 dB while idle.

Cabling requirements

The customer/system integrator must provide the SCSI cable used to
connect the equipment and host computer. It is not included with the
Scanner 9500 or Scanner/Microimager 990.

Electrical and
operating
requirements

This equipment is manufactured to operate within one of the following sets
of power constraints or configured to do so by your Kodak Field Engineer:

Voltage

Hz/Cycles

Amps 9500

Amps 990

90 -110 V ac

50/60 Hz
1 phase

15 amps
@ 100 V ac

15 amps
@ 100 V ac

104 -127 V ac

60 Hz
1 phase

12 amps
@ 120 V ac

12 amps
@ 120 V ac

198 - 254 V ac

50 Hz
1 phase

6 amps
@ 220-240 V ac

10 amps
@ 220-240 V
ac

A separate, dedicated power line with one duplex receptacle is required for
each Scanner.

Two outlets are required:

One individual dedicated branch AC circuit employing one
standard 3-prong, grounded duplex outlets. These outlets are
designated for the machine and service diagnostic equipment.

The outlets should meet the following standards:

Amperage — 15 amp