42 Chapter 5 Maintenance

3

To adequately disinfect the camera, follow the disinfectant
manufacturer’s instructions for the appropriate contact time.

Maintenance

Perform the following maintenance activities.

Daily

Perform these activities daily on the camera and docking station.

Monthly

Wipe the outer covers of the units with a soft, dry cloth.

Table 2

Daily Maintenance

Description

User Instructions and

Precautions

Cautions

Decontamination
and Disinfecting

Prior to disinfectant
remove all visible soil.
Use a wipe to clean the
camera.

Do not rinse.

Allow to dry in the open
air.

Read the Disinfectant
Material Safety Data
Sheet (MSDS) prior to
use.

Do not immerse the
camera in a
disinfecting liquid.

Caution: Do not use disinfectants containing
phenolics or idophors, which will damage the
surface coating.

Caution: Never put the camera in a sterilizing
device or immerse it completely.