GETTING YOUR InDuo™ SYSTEM READY FOR USE

æ OneTouch

®

Ultra Test Strips

Your InDuo™ meter is designed for use only with
OneTouch

®

Ultra Test Strips.

• Store the OneTouch

®

Ultra Test Strip package in a

cool, dry place below 30°C (86ºF). Keep away from
direct sunlight and heat. Do not refrigerate.

• Store your OneTouch

®

Ultra Test Strips in their

original vial only; do not transfer them to a new
bottle or any other container.

• After removing a OneTouch

®

Ultra Test Strip from

the vial, immediately replace the vial cap and close
it tightly.

• With clean, dry hands, you may touch the test

strip anywhere on its surface when removing it
from the vial or inserting it into the meter.

• Use each OneTouch

®

Ultra Test Strip immediately

after removing it from the vial.

• Write the discard date on the vial label when you

open it for the first time. Discard remaining
OneTouch

®

Ultra Test Strips and the vial 3 months

after the vial was first opened.

Contact bars
Insert this end of
OneTouch

®

Ultra

Test Strip into the
InDuo™ meter.

Top edge
Apply a sample here,
where the narrow
channel meets the
top edge of
OneTouch

®

Ultra

Test Strip.

Confirmation
window
This is where you
confirm if a large
enough sample
has been applied
to the top edge.