6.4 Typical fill or drain problems

Fill and drain problems may be due to several causes:

1. There is insufficient reagent

•

Verify that both the reagent bottles and the wax stations have
sufficient reagent.

2. Insufficient pressure or vacuum

The Leica ASP300 is unable to create sufficient vacuum (for a fill) or
pressure (for a drain) in the retort.

•

Verify that the reagent bottle in question is correctly inserted in
the reagent cabinet.

•

Verify that all reagent bottles are fully pushed home into their
sockets.

•

Check whether there is some debris under the retort lid seal
causing a leak.

3. There is a blockage in the air or reagent lines

If there is reagent in the retort that cannot be drained, try to remove
the blockage as follows:

•

Heat the retort to the maximum possible temperature (consistent
with the reagent currently in the retort, if any) and leave it at this
temperature for at least 15 minutes.

•

Whilst the retort is at maximum temperature, attempt to fill or
drain.

•

If the retort is empty:
Run the extended retort clean program. Use the maximum set-
tings for number of clean cycles and temperature (consistent
with the type of reagents/solvents used in the clean program).

If you cannot eliminate the problem by carrying out the above
listed steps, call Leica Technical Service

6.

Trouble shooting