Control solution testing – Lifescan Ultra 2 User Manual

Page 48

Advertising
background image

42

Customer Service

1 800 227-8862

Control solution testing

OneTouch

®

Ultra

®

Control Solution contains

a known amount of glucose and is used to
check that the meter and the test strips are
working properly.

Do a control solution test:

•

to practice the test process instead of
using blood,

•

once a week,

•

whenever you open a new vial of test strips,

•

if you suspect the meter or test strips are not working properly,

•

if you have had repeated unexpected blood glucose results (as described
on pages 28 and 29), or

•

if you drop or damage the meter.

NOTE:
• Use only OneTouch

®

Ultra

®

Control Solution with your

OneTouch

®

Ultra

®

2 Meter.

• Control solution tests must be done at room temperature (68–77°F).

Make sure your meter, test strips, and control solution are at room
temperature before testing.

ƽ

CAUTION:

Do not swallow

control solution; it is not for
human consumption.

Do not apply control solution
to the skin or eyes as it may
cause irritation.

DCO_06344501B_OU2_OB_USe.qxd 12/21/05 8:03 AM Page 42