Wipe the test strip holder cover and base with
a cotton swab or soft cloth moistened with a
10% bleach solution. Be sure to thoroughly
wipe the area on the inside cover. Clean both
sides of the base. Follow with a swab
moistened with water to remove residual
bleach.

▲

CAUTION

Bleach residue left on the test strip holder may
lead to an error message or an inaccurate, high
result.

Dry the test strip holder with a soft cloth or
lint-free tissue. Close the holder and set it aside.

Using a cotton swab or soft cloth moistened
with a 10% bleach solution, wipe the lens area
and contact points. Wipe this area even if it does
not appear to be dirty. Use a swab or cloth
moistened with water to remove residual bleach.
Be careful not to scratch the lens area.

2

3

4

contact points

lens area