IMPORTANT CONTROL SOLUTION TEST INFORMATION

■

Use only SureStep

®

Glucose Control Solution.

■

Check the expiration date. Do not use if expired.

■

Record the discard date on the control solution vial. Discard the vial
three months after first opening it.

■

The control range printed on the test strip bottle is for SureStep

Glucose Control Solution only. It is NOT a recommended range for your
blood glucose level.

■

Do not store the glucose control solution at room temperatures above
30°C (86°F). Do not refrigerate or freeze.

■

If a control solution test falls outside of the control range, do not use
the meter to test your blood until you get a control solution test result
within the range. Call the LifeScan Customer Care Line at 1 800 663-5521
(U.S.A. 1 800 227-8862) if you need assistance.

WARNINGS

■

Control solution is NOT TO BE USED as a cleaning solution.

■

DO NOT swallow or inject control solution or put it in your eyes.

■

Control solution contains a dye that stains clothing.