Blender

Air-Oxygen

1. Disconnect all gas connections and equipment before

cleaning.

2. Clean exterior surfaces with a cloth dampened with mild

detergent and water.

3. Wipe dry with a clean cloth.

MAiNTENANCE

The following maintenance on the Air-Oxygen Blender must be
performed by a trained service technician:

Every month verify proper functioning of alarm.
Every year conduct the Operational Verification Procedure (OVP).
Every 2 years the Air-Oxygen Blender should be serviced.

PM5200 (P/N 505407)

PM5300 (P/N 504932)

Refer to the Air-Oxygen Blender Service manual (P/N 504827)
for complete details regarding further maintenance and
testing.

TECHNiCAL dESCRiPTiON

For a complete Technical Description of the Air-Oxygen Blender
and list of Replacement Parts, reference the Air-Oxygen Blender
Service manual (P/N 504827) available on the Internet;
www.precisionmedical.com.

RETURNS

Returned products require a Returned Goods Authorization
(RGA) number, contact Precision medical, Inc. All returns must
be packaged in sealed containers to prevent damage. Precision
medical, Inc. will not be responsible for goods damaged in transit.
Refer to Precision medical, Inc. Return Policy available on the
Internet; www.precisionmedical.com.