Controls and Connectors
Monitor Installation and Patient Safety 11-25
M
onitor
Ins
ta
lla
ti
on
a
nd Pa
ti
e
n
t Sa
fe
ty
M1094B Display Module
Controls of the
M1092A,
M1094A and
M1094B
Displays
1. Video Termination Switch; thus switch is used to terminate the
video signal at this display: this must be switched to the “75
Ω
”
position if only a master display is used and must be switched to the
“HIGH IMP.” position if a second, duplicate display is used.
Maximum Voltages
System Power Connector
60 V
Human Interface Connector
12 V
Local Power Connector
240 V
Video Out Connector
5 V
Video In Connector
5 V