Oxycrg alarm recording, Oxycrg alarm recording -38 – Philips V24CT User Manual
Page 210
 
oxyCRG Recordings
6-38 Recording Functions
Re
c
o
rd
ing
Func
tions
oxyCRG 
Alarm 
Recording
Pressing the
softkey allows you to choose
between “Standard” alarm recording or “oxyCRG” alarm recording.
If you select “oxyCRG”, an oxyCRG recording is started when an alarm 
occurs. Six minutes of pre-alarm data and two minutes of post-alarm 
data are recorded. Any manually initiated recording that is in progress 
will be interrupted, and the oxyCRG recorded instead. If a second alarm 
occurs while an oxyCRG alarm recording is already in progress, the 
recording will be extended to two minutes after the second alarm 
appears on the strip.
If you select “Standard”, your selected realtime waves are recorded 
when an alarm occurs, as described earlier in this chapter.