V I I . C A R E A N D M A I N T E N A N C E

To replace the batteries:

1. Turn off the power to the controller.

2. Make sure that the power chair is in drive mode. See III. “Your Power Chair.”

3. Remove the foot platform by squeezing the release handles (see figure 27) and lifting the foot platform bracket

up and away from the power base.

4. Disconnect the quick release battery connectors. See figure 7.

5. Remove the batteries.

6. Disconnect the battery harness from each battery.

7. Connect the battery harness to the new front battery according to the battery wiring diagram. See figure 27.

8. Connect the battery harness to the new rear battery according to the battery wiring diagram. See figure 27.

WARNING! Make sure you tighten the fasteners so that the connections are secure.

9. Install the new batteries.

10. Reinstall the front cover and foot platform.

11. Charge the batteries. See VI. “Batteries and Charging.”

When to See Your Quantum Rehab Provider for Service

The following symptoms could indicate a serious problem with your power chair. If necessary, contact your

Quantum Rehab Provider. When calling, have the model number, serial number, nature of the problem, and the

error code if available.

Motor noise

Frayed harnesses

Cracked or broken connectors

Uneven wear on any of the tires

Jerky motion

Pulling to one side

Bent or broken wheel assemblies

Does not power up

Powers up, but does not move

Loose seat or seating components

Corrective Maintenance

If the battery condition meter does not light up when you turn on the power:

Check the harness connections. Make sure they are tight.

Check the circuit breaker. Reset it if necessary.

Check the battery connections.

If the above conditions prove normal, you can load test the batteries with a battery load tester. These testers are

available at automotive parts stores. Disconnect both batteries before load testing and follow the directions that

come with the load tester. If either one of the batteries fails the load test, replace both of them. If your power chair

still does not power up, contact your Quantum Rehab Provider.