Multiple monitors – National Instruments FPT-1015 User Manual

Page 25

Advertising
background image

Appendix C

Touchscreen

FPT-1015 User Manual

C-2

ni.com

Standard Calibration

When you click this button, arrows
appear pointing to red squares. Use your
finger or stylus to touch the red squares in
sequence. After you touch the fifth red
square, calibration is complete. To skip,
press <Esc>.

Note

To access the Advanced Calibration option, go to the Tools tab and select Advanced

Calibration.

Advanced Calibration

Advanced calibration uses 4, 9, 16, or
25 points to effectively calibrate the
linearity of aged touchscreens. Click
this button and touch the red squares in
sequence with a stylus. To skip, press
<Esc>.

Note

The older the touchscreen, the more advanced calibration points you need for

accurate calibration. Use a stylus during advanced calibration for greater accuracy.

Plot Calibration Data

When you select this function, a touch
panel linearity comparison graph appears
after you finish advanced calibration.
The blue lines show linearity before
calibration, and black lines show linearity
after calibration.

Multiple Monitors

The Multiple Monitors function supports two to six touchscreen displays
for one system. Each monitor requires its own PenMount touchscreen
control board, installed inside the display or in a central unit. You must
connect the control boards to the computer COM ports via the RS232
interface. Driver installation procedures are the same as for a single
monitor. Multiple Monitors supports the following modes:

Windows Extend Monitor function

Matrox DualHead multiscreen function

nVidia nView function

Note

The Multiple Monitors function is for multiple displays only. Do not use this

function with only one touchscreen display.

Advertising