Part 3 - trouble shooting – Omega FD6003 User Manual

Page 36

Advertising
background image

3-20

PART 3 - TROUBLE SHOOTING

CONDITION

POSSIBLE CAUSE

Unit does not turn “ON” when power is
applied

•

Verify that AC power source is live.

•

Test the fuse

•

Verify that proper power supply jumpers
are in place.

FAULT CONDITION is indicated on LCD
indicator

•

Fault conditions can arise from several
causes: electrical surges, short circuits,
open circuits, etc. To clear a “Fault Condi-
tion”, press 2nd Function then press Re-
set
. Use the arrow keys to change the
Fault code number to “0”. Press Enter. If
the meter resumes measuring flow, no
permanent damage was incurred.

•

Ensure that the transducer is properly
mounted to the pipe.

•

Verify that transducer connections are
correct

•

Ensure that the pipe is full of moving
liquid.

•

If SIGNAL STR is less than 000100 counts
and flow rate is greater than 1 FPS [0.3
MPS], adjust GAIN control (R13 on the
Main PCB) till SIGNAL STR reaches at
least 000125 counts.

•

On cleaner liquids, move the transducers
closer to a 90

o

pipe elbow.

•

If GND connection and pipe are at
different potentials, ground FD6000 to
pipe potential.

•

If Variable Frequency Drives are being
utilized, verify that the FD6000 obtains a
flow indication when the pump turns OFF.
If it does, contact OMEGA Engineering.