Omega Speaker Systems RH/Temperature Transmitter HX93A User Manual

Page 17

Advertising
background image

13

16. Maintenance

If your Humidity transmitter will be used in a dusty environment, the
protective sensor filter, if clogged, may be removed for cleaning.
Unscrew the protective cover and gently blow compressed air through
the filter screen. A soft brush may also be used to remove dirt particles
from the screen.
If the sensor is subjected to 100% condensation, it must be dried to
obtain correct readings. There will be no permanent damage or
calibration shift to the unit.
Units should not be exposed to high concentrations of ammonia or
alcohol vapors.

17. Specifications

Relative Humidity

Measuring Range:

3 - 95% (non-condensing)

Accuracy:

±2.5% @ 72ºF with an added temperature
coefficient error of -.06% RH/ºF (both
increasing and decreasing in temperature).

Repeatability:

± 1 % RH

Operating
Temperature Range:

-20 to 75ºC (-4 to 167ºF)

Output:

Model: HX93AC

4 to 20 mA (Scaled for 0 to 100% RH)

Model : HX93AV

0 to 1 Vdc (Scaled for 0 to 100% RH)

Power:

6 –30 Vdc @ 20mA

Max Loop Resistance:

Ohms = (V supply – 6 V)/.02 A

RH Time Constant (90%
response at 25ºC, in
moving air at 1m/s):

>10 seconds, 10 to 90% RH
>15 seconds, 90 to 10% RH

Sensor Type:

Thin Film Polymer Capacitor

Temperature

Measuring Range:

-20 to 75ºC (-4 to 167ºF)

Accuracy:

± 0.6ºC (1ºF)

Repeatability:

± 0.3ºC (0.5ºF)

Output:

Model: HX93AC

4 to 20 mA (Scaled for 0 to 100% RH)

Model : HX93AV

0 to 1 Vdc (Scaled for 0 to 100% RH)

Power:

No Displays

6 -30 Vdc @ 20mA

With Displays

12 - 30 Vdc @ 30mA

16

HX93A, HX93DA SERIES

RH/Temperature Transmitter

Advertising
This manual is related to the following products: