Installation of solar radiation sensor – Omega WMS-16 User Manual

Page 18

Advertising
background image

17

The auxiliary temperature sensor is a thermistor attached to a two conductor cable. Connect one conductor to
A5 terminal and the other conductor to a GROUND terminal.

INITIALIZATION OF THE AUXILIARY TEMPERATURE SENSOR.

Press EXC to get the MAIN MENU.

From the MAIN MENU press number 1. STATION SET UP.

From the Station Set Up menu press number 6. ADD OR REMOVE SENSORS. Then press Y, then ENTER.

From the Sensor Type Menu select number 4. Temperature. Then press
ENTER.

From the Temperature Sensors Menu select number 1. (WMS-16T). Then press ENTER.

The menu then asks: Enter temperature offset or press Q to quit. Press Q.

From the Sensor Input Channel Menu select number A6. Then press ENTER.

If you intend to add another sensor select A.
If you do not want to add another sensor select Q.
You will be asked if you are sure that you want to erase memory. If you have added or deleted a sensor select Y
if not select N and press ENTER.

Installation of Solar Radiation Sensor

Connect the solar radiation sensor to the terminal strip on the back of the control module. This sensor has a two
wire cable. Attach the WHITE conductor to the A1 terminal and the BLACK conductor to a GROUND
terminal.
INITIALIZATION OF THE SOLAR RADIATION SENSOR

Press EXC to get the MAIN MENU.

From the MAIN MENU press number 1. STATION SETUP.

From the Station Set Up menu press number 6. ADD OR REMOVE SENSORS. Then press Y, then ENTER.

From the Sensor Type menu select number 8. Solar Radiation. Then press ENTER.

From the Solar Radiation menu select number 1. Model 240-140. Then press ENTER.

Next enter the calibration factor from the calibration sheet, which was supplied with the sensor.

Then press ENTER.