Calibrating, Calibrating outputs – Quantum Data 700 Series User Manual

Page 10

Advertising
background image

10

Calibrating

Calibrating Outputs

1.

Select Options > Calibrate > Reset to set the generator’s
calibration factors to default nominal values.

2.

Press Options > Calibrate > Full Scale.

3.

Set the voltmeter’s scale factor to be able to measure a nominal
1000 millivolts DC.

4.

Using a VGA-to-BNC cable, connect the generator’s red output to
the positive input on the voltmeter, with a 75 ohm (+/- 1%) input
terminator. Connect the negative lead of the meter to any ground
pin.

5.

Select R_FS+ or -R_FS to set the DC output level to 1000 mV
+/- 3 mV.

6.

Repeat steps 4 and 5 for the green and blue outputs.

7.

Select Save.

8.

Select Options > Calibrate > Zero.

9.

Connect the red output to the positive input of the voltmeter.

10. Select R_Zero+ or -R_Zero to set the DC output level to 0 mV +/- 3 mV.

11. Repeat steps 9 and 10 for the green and blue components.

12. Select Save.

13. Select Options > NTSC_CAL.

14. Adjust the oscilloscope's timebase and sweep trigger to view the entire waveform. Adjust the

G_NTSC so that its chrominance is 100 IRE or 714mV.

15. Using a vector scope, use the R_NTSC and B_NTSC options to adjust the red and green outputs

using so that their signals align with the 2% boxes.

16. Press Save.

Note: The modulated RF output does not have any calibration settings.

_____

Calibrate

_

_

Full

_

Scale

____

Zero

_

Reset

_

R_FS+

_____ ___

-R_FS

_

G_FS+

_______ _

-G_FS

_

B_FS+

_______ _

-B_FS

_

Save

_________ _

Back

_

R_Zero+

____

-R_Zero

_

G_Zero+

____

-G_Zero

_

B_Zero+

____

-B_Zero

_

Save

__________

Back

Advertising