Toshiba Electromagnetic Flowmeter LF470/LF612 User Manual

Page 2

Advertising
background image

TIC-LF470B

2

Model LF470 Detectors

Fluid pressure: -15 to 150 psi, or -1.0 to 10 bar

(-0.1 to 1 MPa)

Principal materials:

Measurement tube – Alumina ceramics
Electrodes – Platinum
Pipe connection port:

Standard – 316 stainless steel

Option – Ti (titanium),

Polyvinyl chloride (shock-resistant)+
Ta (Tantalum) for grounding plate,
Polyvinyl chloride (shock-resistant)+
Pt-Ir grounding plate,
Hastelloy C (Equivalent)

Dedicated preformed cables:

・

Signal cable

– 2-wire shielded chloroprene

sheathed cable

Overall length: 0.28 inch (7mm)
Length: 0.20 inch (5m)

Excitation cable – 3-wiren chloroprene sheathed

cable

Overall length: 0.28 inch (7mm)

Length:

0.20 inch (5m)

Coating: phthalic acid resin coating pearl-gray

colored

Weight: approximately 2.2 lb (1.0 kg)

(for each size excluding cables)

Model LF612 converters

Input signals

Analog signal — the voltage signal from detector,

proportional to process flow rate (for LF612

separate type converter).

Digital input DI (opt.)

Signal type: 20 to 30Vdc voltage signal

Input resistance: 2.7k

Ω

Number of inputs: one point

DI function — One of the following functions can be

assigned to the optional DI signal.

Range switching — Selects either the higher or lower

range in the unidirectional or bidirectional 2-range

setting.

Totalizer control — Starts and stops the built-in

totalizer.

Fixed-value outputs —Outputs fixed-values for

current and pulse outputs.

Zero adjustment — Executes zero adjustment

(on-stream at zero flow rate).

Output signals

Current output:

4–20mAdc (load resistance 0 to 750

Ω)

Note: The current output cannot be used with the

PROFIBUS-PA communication.

Digital outputs — One point (std.) and one more

point is optionally available as follows.

Digital output DO1 (std.):

Output type: Transistor open collector
Number of outputs: One point
Output capacity: 30Vdc, 200mA maximum

Digital output DO2 (opt.):

Output type:

Solid-state relay output (non polarity)

Number of outputs: One point
Output capacity: 150Vdc, 150mA maximum

or 150Vac (peak to peak), 100mA maximum

DO1 and DO2 functions — One of the following

functions can be assigned to DO1 (std.) and/or
DO2 (opt.)

• Pulse output (available only for DO1, DO2)

Pulse rate: 3.6 to 36,000,000 pulses/hr (DO1)

3.6 to 360,000 pulses/hr (DO2)

(Over 3,600,000 pulses/hr, auto-setting)

Pulse width: 0.5 to 500ms (but less than half of
the period for 100% flow rate)

Note: The same and simultaneous pulse is not

available between DO1 and DO2.)

• Multi-range selection outputs (Note 1)
• High, High high, Low, and/or Low low alarm

outputs (Note 2)

• Empty pipe alarm output
• Digital Output Active Status (DO1 and DO2)

(Note 2)

• Preset count output
• Converter failure alarm output

Note 1: Two outputs (DO1 and DO2) are needed for

4-range switching and forward/reverse 2-range
switching.

Note 2:Normal Open (default set) or Normal Close is

selected for alarm outputs when programming.

When power failure occurs, unit will be fault to
Normal Open.