Installation 3

Installation 3

Installation 3

Installation 3

Installation 3

3-3

Teledyne Analytical Instruments

The analyzer should have a vent line of ¼" diameter tubing at least two

feet long, running downward

downward

downward

downward

downward from the vent connection. This is to prevent

air from diffusing into the reservoir and dissolving into the humidifier make-
up water.

If it is not desirable to vent the sample into the atmosphere, a vent line

to carry the sample to a suitable venting area will be required. The sample
leaves the vent connection of the analyzer saturated with water vapor at a
temperature somewhat above ambient, so a suitable trap to remove conden-
sate without plugging the vent line will be required. The vent line should
also be arranged so that it cannot become plugged by dirt or dust.

Figure 2-2: General Connection Diagram

See the specific Interconnection Diagram for your instrument in the drawing
package located at the back of the manual. See also any Addenda that may be
included with this manual.