Quickstart Guide

Trace Oxygen Analyzer


Teledyne Analytical Instruments

14

•

Press the Alarm button on the front panel to enter the Alarm function. Make sure that

AL–1 is blinking.

AL—1 AL—2

Choose Alarm

•

Using the ◄► arrow keys to move the blinking over to AL–1. Then press Enter.

AL—1 1000 ppm HI

Dft—N Fs—N Ltch—N

Five parameters can be changed on this screen:

•

Value of the alarm setpoint,

AL–1 ####

ppm (oxygen); value can be set from 0.00

to 10,000.00 ppm

•

Out-of-range direction,

HI

or

LO

•

Defeated?

Dft–Y/N

(Yes/No)

•

Failsafe?

Fs–Y/N

(Yes/No)

•

Latching?

Ltch–Y/N

(Yes/No).

•

To change any of the alarm parameters, use the ◄► arrow keys to move the blinking
over to the specific parameter.

•

Use the ▲▼ arrow keys to change the value of the parameter. Holding down the key
speeds up the incrementing or decrementing.

•

Once the parameters for alarm 1 have been set, press Alarms again, and repeat this
procedure for alarm 2 (

AL–2

).

•

To reset a latched alarm, go to

Dft–

and then press either ▲ two times or ▼ two

times. (Toggle it to Y and then back to N.)

—or—

Go to

Ltch–

and then press either ▲ two times or ▼ two times. (Toggle it to N and

back to Y.)

Set The Range Function

To set the 3 programmable ranges:

•

Enter the range function mode by pressing the Range button on the front panel.

L—1 M—10

H—100 Mode—AUTO