BioOptix™ 10

Appendix B Fluid Path Diagrams

This appendix contains simplified diagrams of the plumbing and
valves internal to the BioOptix 10 system. Various operating
states are depicted for a single channel. A complete plumbing
diagram can be found in Figure B-7.

An understanding of fluids paths can be helpful when designing
custom template files (Appendix A), manually controlling the
system, or troubleshooting.

The valves in the diagrams are:

• Buffer–This valve switches between Buffer A and

Buffer B to produce the correct mix (%B) of buffer
solution.

• Sample–This valve allows the pump to draw either

sample from the Sample In ports or buffer solution.

• Channel–This valve controls allows fluid to be pumped

only when the channel is active.

• Mode–This manually-controlled front-panel valve has

three positions:

❍

Operate–passes fluid from the pump to the column.

❍

Manual Inject–directs injected fluid onto the column.

❍

Purge Pump–diverts fluid from the pump to a purge
port that is common to all channels.

• Effluent–This valve sends the effluent to the Foxy

fraction collector or to the channel’s waste port.

B.1 Fluid-type Blocks

Fluid blocks pump buffer solution through the column (Figure
B-1).

Figure B-1 Fluid path for Fluid-type blocks

Buffer

Sample

Channel

Mode

Effluent

Pump

To Waste

To Foxy

Sample In

Buffer A

Buffer B

To Pump Purge

Manual Sample Injection

Column