TW-357 REV E



8

KRYOS Specifications
Configurations:

Designed

exclusively for all Taylor-Wharton LABS CryoStorage

Systems

Power Supply:

16.5 VAC, 40 Voltamps – Standard Battery Backup Version
24 VAC, 40 Voltamps (No Battery)
(For power outlets other than North American 110 VAC contact
Taylor-Wharton Customer Service.)

Sensor Assembly:

Freeze-Guard Assembly – Standard
8-Thermistor Assembly– Optional
4-Thermistor Assembly – Optional

Thermocouples:

Operates with none, 1 or 2 Type T
Thermocouples (1 piece standard)


Solenoid

Valve:

12

VDC

cryogenic solenoid valve – Standard

Control Type:

Liquid Level Control or Liquid Level Control with Temperature
Control

Security:

User Access via 2-digit password
Power On/Off Password
Menu access Password

Alarms:

Audible

Visual red LED
Alarm description
Remote alarm relay

Diagnostics:

Circuit diagnostics at start-up
Sensor diagnostics from front panel
Thermocouple diagnostics from front panel
Manual Test for audible, visual and remote alarms

Temp. Calibration:

Automated calibration from the front panel


Communications:

RS-232 Serial Port for 2-way communications capable


Logging Capacity:

System Logs (4096 events)
Alarm Logs (4096 events)
Temperature Logs (32,768 events)

Battery:

A CR2032 coin cell battery is used to back up time/date