Xa positioner module, Curve module, Voi lut module – Toshiba XIDF-038A User Manual
Page 23: Sop common module, 14 curve module, 15 voi lut module, 16 sop common module

No. MIIXR0002EAC
8.5.13 XA Positioner Module
Table 28
Attribute Name
Tag
Type
Attribute Description
Distance Source to Detector
(0018, 1110)
3
Always set [mm]
Estimated Radiographic Magnification
Factor
(0018,1114) 3
Always
set
Positioner Motion
(0018, 1500)
2C
Always set ("STATIC")
Positioner Primary Angle
(0018, 1510)
2
Always set [degree]
Positioner Secondary Angle
(0018, 1511)
2
Always set [degree]
8.5.14 Curve
Module
Table 29
Attribute Name
Tag
Type
Attribute Description
Curve Dimensions
(5000, 0005)
1
Always set
Number of Points
(5000, 0010)
1
Always set
Type of Data
(5000, 0020)
1
Always set
Axis Units
(5000, 0030)
3
Always set
Data Value Representation
(5000, 0103)
1
Always set
Curve Data Descriptor
(5000, 0110)
1C
Always set
Coordinate Start Value
(5000, 0112)
1C
Always set
Coordinate Step Value
(5000, 0114)
1C
Always set
Curve Data
(5000, 3000)
1
Always set
Note : This module is set only if ECG data is available.
8.5.15 VOI LUT Module
Table 30
Attribute Name
Tag
Type
Attribute Description
Window Center
(0028, 1050)
3
Always set
Window Width
(0028, 1051)
1C
Always set
8.5.16 SOP Common Module
Table 31
Attribute Name
Tag
Type
Attribute Description
SOP Class UID
(0008, 0016)
1
Always set
SOP Instance UID
(0008, 0018)
1
Always set
E