15 technical specifications – Topcom BPM Wrist 7500 User Manual

Page 11

Advertising
background image

11

TOPCOM BPM Wrist 7500

ENGLISH

To erase all data from one memory zone:

•

Press the START/STOP button to activate the display.

•

Press the UP or DOWN button to select the desired
memory zone.

•

Press the MEMORY button to see the average of all
measurement results.

•

Press the MODE and the DOWN button simultaneously to
erase all data from the selected memory zone.

15 Technical specifications

*Specifications are subject to change without notice.

Power source

Measurement method
Measurement zone

Accuracy

Pressure sensor
Inflation
Deflation
Memory capacity
Auto-shut-off
Operation environment

Storage environment

Weight
Wrist circumference
Limited patient
Dimensions
K.B.

Classification

3V DC: Two LR03 (AAA) Non rechargeable batteries
Oscillometric
Pressure: 0 ~ 300 mmHg
Pulse: 40 ~ 199 beats/minute
Pressure: ± 3 mmHg
Pulse: within ± 5% of reading
Semi-conductor
Pump driven
Automatic Pressure release valve
3 Memory zones for each 40 memories
1 minute after last key operation
+10°C ~ + 40°C

85% RH

-20°C ~ + 70°C

85% RH

148g (including batteries)
135 ~ 195 mm
Over 18 years old
74(L) x 84(W) x 32(H) mm
Short time operation 2 minutes

• Internally powered equipment
• Classification of type BF (applied part)
• IPX0
• Not suitable for use in presence of flammable anaes-

thetic mixture with air or with oxygen or nitrous oxide

• Continuous operation with short-time loading

≤

≤

The CE symbol indicates that the unit complies with the

essential requirements of the directive 93/42/EEC.