TELEDYNE ANALYTICAL INSTRUMENTS

3.6.3 RS-232 Serial Digital Port

The RS-232 port is configured as data terminal equipment (DTE) and uses a 9-pin

D connector wired as follows:

1.

No connection

2.

Transmit Data

3.

Receive Data

4.

N/C

5.

Ground

6.

N/C

7.

RTC

8.

CTS

9.

N/C

Connecting the analyzer to a modem requires a PC to modem cable. When con-

necting the analyzer directly to a PC, terminal or other data terminal equipment, a
“null-modem” cable is required.

The following serial interface default parameters are used:

Defaults

Programmable Options

1200 Baud

1200 Baud

8 Bits

5, 6, 7, or 8 Bits

No Parity

Even, Odd, or None

1 Stop Bit

1 or 2 Stop Bits

3.7

Installation Checklist

Have you:

•

Checked for leaks from any of the rear panel connections?

•

Checked the sample and span gas pressures (5–10 psi)?

•

Checked the instrument air pressure (70–80 psi)?