Getting Started

1Ć6

MTG100 & MTG300 MPEG Generator User Manual

Installation

Before you install the instrument, refer to the General Safety Summary at the
front of this manual for power source, grounding, and other safety information.

Verify that the location of your installation has the proper operating environment.

CAUTION. Damage to the instrument can occur if this instrument is powered on
at temperatures outside the specified temperature range.

The MPEG generators operate correctly in ambient temperatures from +5

_ C to

+40

_ C and in relative humidity from 20% to 80%. For complete operating

environment information, refer to Appendix A: Specifications.

Verify that there is at least 5.0 cm (2 inches) of clearance on top and bottom,
15.0 cm (6 inches) on the left and right sides, and 7.5 cm (3 inches) at the rear of
the instrument to allow for heat dissipation. Verify that the air intake holes on the
sides and bottom of the cabinet are not obstructed.

NOTE. If you are installing the instrument in a dedicated rack, refer to the
instruction sheet that comes with the rackmounting kit for proper installation
procedures.

When the generator is mounted in a 19-inch rack, verify that there is at least one
unit of clearance above the MPEG generator.

Check that your location provides the proper electrical power requirements as
listed in Table 1-5.

Table 1Ć5: AC line power requirements

Parameter

Description

Line Voltage Range

90 V - 250 V

Line frequency

48 Hz - 63 Hz

Maximum power

200 W

Environmental Operating

Requirements

AC Line Voltage

Requirements

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com