Troubleshooting, Specifications, Page check points problem – Tanita BD-815U User Manual

Page 14

Advertising
background image

GB

GB

27

26

Troubleshooting

Page

Check points

Problem

If you are concerned that the scale may not be functioning correctly,
please check the following points before requesting repairs.

• Have the batteries been removed, or has the AC

adapter been unplugged?

• Are the batteries running low?
• Is anything stuck to the terminals of the batter-

ies?

s

P.12

Nothing is displayed

when the

key is

pressed.

• The batteries are running out. Please replace

them immediately or use the AC adapter.

s

P.12

is displayed.

• Is any pressure being applied to the baby basket

when the power is turned on?

• Has basket been removed?

s

P.14

After the power is
turned on,
is displayed, and an
alarm sounds.

• The maximum measurable weight capacity for

the scale (33lb / 15 kg) is exceeded.

s

P.7

is shown

during measure-
ment.

Specifications

Power supply

Model

BD-815U

Capacity

33 lb (15000g)

d (division)

0.1oz(0-12lb)

0.2oz(12-33lb)

2g (0-6000g)

5g (6000-15000g)

Maximum tare

4lb (2000g)

Extended calibration (uncertified)

0.05oz (1g)

Display panel

Backlit LCD
(Backlight is on only when the AC adapter is used)

External interface

RS-232C (D-sub 9-pin female connector),
unidirectional output

Input: AC 120 V (60 Hz) Output: DC 9 V / 200mA

Power consumption

0.12 W or less (using the batteries)
0.35 W or less (using the AC adapter)

Operating temperature

41˚F - 95˚F (5˚C - 35˚C)

Overall dimensions

D16.1 W23.6 H6.2in (D409 W600 H157mm)

Main unit mass

Approx. 10.1lb(4.6kg)

Main material

ABS resin

Made in

Japan

Regular voltage

DC 6 V (using the batteries)
DC 9 V (using the AC adapter)

Consumption current

20mA (using the batteries)
38mA (using the AC adapter)

Storage temperature

14˚F - 122˚F (-10˚C - 50˚C)

Advertising