Chapter 3

Operating Valves Automatically

Operation

Operating Manual

39

2. Scroll to

Add Run Time Event

and press ENTER.

3. Select

Valve

and press ENTER.

4. Scroll to the sampling valve that you want to add and press ENTER.

The following screen appears:

The

Inject at

parameter is the time into the run that the sample is to be

injected. The

Inject for

parameter is the time that the valve remains into the

inject position. After that time, the valve returns to the Load position. If the

Inject for

time exceeds the GC run time, the GC will automatically reset the

valve to the Load position at the end of the run.

5. Repeat as necessary for any remaining sampling valves. If you need to edit a

run table entry, select the entry to be edited and press ENTER.

NOTE

To delete a run table entry, select it and press CLEAR. The TRACE GC Ultra will ask you if
you really want to delete the entry. Press YES.

SELECT EVENT to add

Signal
Valve

<

External Event

SELECT PARAM to Add

Valve #1

Sampling

<

Valve #2

Switching

Valve #3

Sampling

RUN TIME EVENT

Valve #1

Sampling

Inject at

0.00

Inject for

0.00