VXI SM8000 User Manual

Page 59

Advertising
background image

VXI Technology, Inc.

SM8000 Series Programming

59

Note

To operate the modules correctly, the SM8000 must be loaded with a valid Address in the Address

Register. The SM8000 is hard coded at the factory with the optical modules default address of

73 => 49h, and may be used to generate the address used in the command. This is the address of all

attenuators as shipped from the factory. During the programming of the optical module, the

programmer may wish to omit sending the module’s address over the VXI Bus, letting the SM8000

generate the default address that is used in the command string. This could possibly increase

throughput, by decreasing VXI Bus traffic, if the modules are receiving many commands, although this

is only true if the optical module’s address is not changed by the user. If the address is to be changed,

IT IS IMPERATIVE THAT THE NEW ADDRESS BE WRITTEN DOWN. Failure to do so will result

in an inability to control the module. All four possible modules may have the same address. The

SM8000 controls them on separate internal busses.


VTI STRONGLY RECOMMENDS THAT THE OPTICAL MODULE’S ADDRESS BE LEFT AT 49h.