Welch Allyn OAE Hearing Screener Directions for Use 705353 Rev E

17

Quick Guide: Causes of ‘Refer”

r

Welch Allyn OAE Hearing Screener

Troubleshooting Quick Guide: Causes of Refer Results

Noise

Problem

Possible solutions

♦

Noisy surroundings

— Other people in or near test site.

— Devices such as fans, air

conditioners, intercoms, and
public address systems.

♦

Eliminate or reduce noise

— Isolate and completely sound-proof the test site (optimal).
— Ask others in or near test site to refrain from talking or other

activity.

— Ask noisy people to move to another location; ask parents to

move noisy children to another location.

—

Deactivate or remove any device that is a potential source of
noise.

—

Use larger probe tip, if possible.

♦

Patient noise

— Crying or struggling against the

probe

—

Sucking a pacifier or bottle

—

Excessive movement

—

Talking

♦

FOR NEWBORNS OR INFANTS:

— Swaddle child in blanket and test while child is asleep.
— Ask parent to hold child comfortably and to prevent child from

pulling probe or cable.

♦

FOR TODDLERS:

— Distract fussy toddlers with a quiet toy or book.
— Ask parent to help by holding the toddler comfortably on lap or

shoulder.

♦

FOR OLDER CHILDREN:

— Explain that being quiet will speed up the test and reduce the

possibility of retesting.

Middle Ear Involvement

Problem

Possible solutions

♦

Middle ear conditions that add

mass or stiffness to the middle ear
such as:

— Otitis media

— Severe Eustachian tube

disfunction

— Otosclerosis

♦

Refer results caused by temporary conditions that cause middle

ear involvement will improve as the middle ear returns to normal.

Advertising