TILT-IN-SPACE GAS SPRING TROUBLESHOOTING

PROBLEM:
If the chair actuates without squeezing the release

lever, the cable assembly needs to be adjusted.

CORRECTION:
1. Lean the chair on its side and remove clevis

pin and wire rue ring that secure the top of

the gas spring to the frame. Do not remove

the cable.

2. Loosen the nut above the gas spring bracket.

Rotate the gas spring out of the bracket

(counterclockwise) 1/2 turn. Retighten nut on

bracket, and reassemble the gas spring back

to the frame. Check that cable is snapped into

bracket (See fig. 14).

3. Test the operation of chair. If the chair still

actuates without squeezing release lever,

repeat the above procedure.

PROBLEM:
If the chair does not actuate with a person in the

chair while squeezing the release lever and

pushing on the back, then the cable is too loose

and the assembly needs to be adjusted.

CORRECTION:
1. Ensure the cable is properly engaged onto the

bracket assembly. If cable is not engaged on

bracket, engage as shown in figure 15.

2. With a person in the chair, recheck cable.

If problem persists, go to step 3.

3. Lean the chair on its side and remove clevis

pin and wire ring that secure the top of the gas

spring to the frame. Do not remove the cable.

4. Loosen the nut above the gas spring bracket.

Rotate the gas spring into the bracket

(clockwise) 1/2 turn. Retighten nut on bracket

and reassemble the gas spring back to the

frame (See fig. 14).

5. Test the operation of chair. If the chair does

not actuate with a person in the chair while

squeezing the release lever, repeat steps 3-5.

Gas Spring Cylinder

Bracket Assembly

Cable

Nut

In

Out

Rue Ring

Clevis Pin

Engage as shown

Cable

Bracket Assembly

Figure 14

Figure 15