Storage
Your Saunders Lumbar HomeTrac should be stored in its protective
case. Replace the pump in the protective foam enclosure. Fold all
belts, closing buckles where applicable. Your Saunders STx should be
stored in the closed or folded position with all components stored in
the storage compartment.

Troubleshooting
Saunders Lumbar Traction Devices feature a very accurate, sensitive
gauge. The pump is speci cally designed not to leak in the “HOLD”
position. If you notice a drop in the gauge reading when in the “HOLD”
position it is probably because you have relaxed or moved slightly. If
this happens, return to the “PUMP” position and pump back to the
desired force, then reset to “HOLD”. If the traction device loses air rapidly,
there may be a leak in the system.

Questions to Ask:

Is the pump in the “PUMP” position when pumping?

• Is the pump in the “HOLD” position when receiving treatment?

Are you lying still? (Any movement will cause ctuation).

Are you pumping slowly?

Are all the connections tight (i.e. hose cap, lens cap)

If the unit is leaking air, try the following:

Check the air hose connections at the hand pump

and gently tighten as necessary. Do not over-tighten as the
threads may strip.

Tighten the hand pump shaft sleeve cap located above the

serial number label.

active

forever

.com

www.ActiveForever.com

Advertising