CAUTION The Inogen One® G3 battery acts as a secondary power supply in the event

of a planned or unexpected loss of the AC or DC external power supply. When

operating the Inogen One® G3 from an AC or DC external power supply, a

properly inserted Inogen One® G3 battery should be maintained in the unit.

This procedure will ensure uninterrupted operation and will operate all

alarms and alerts in the event of a loss of the external power supply.

4. Connect the Power Supply.

Connect the AC input plug to the power supply. Connect the AC power

plug to the power source and connect the power output plug to the

Inogen One® G3. The green LED on the power supply will be illuminated

and a beep will sound from the concentrator.

CAUTION Ensure the power supply is in a well ventilated location as it relies on

air circulation for heat dissipation. The power supply may become hot during

operation. Make sure the power supply cools down before handling.

CAUTION The power supply is not water resistant.

CAUTION Do not disassemble the power supply. This may lead to component failure

and/or safety risk.

CAUTION Do not place anything in the power supply port other than the supplied wall

cord. Avoid the use of electrical extension cords with the Inogen One® G3.

If an extension cord must be used, use an extension cord that has an

Underwriters Laboratory (UL) Mark and a minimum wire thickness of

18 gauge. Do not connect any other devices to the same extension cord.

NOTE

Under certain conditions (see Technical Specifications) the power

supply may shut down. The green LED will blink or will no longer be

illuminated. If this occurs, disconnect the power supply for at least 10

seconds and reconnect.