Setup – Activeforever OptiFlex 3 Knee CPM (OptiFlex3 Knee CPM) User Manual
Page 18
 
OptiFlex® 3 Continuous Passive Motion (CPM) Therapy Unit
15
SETUP
NEW PATIENT SETUP
Securing Patient
After patient is positioned and all 
adjustments are made, secure the patient 
to the OptiFlex 3 using the strap included 
in the Softgoods Kit.
Starting/Stopping Treatment
After all parameters have been made, see 
pages 7 through 11, start treatment by 
pressing the Start/Stop (Emergency Stop) 
button.
Stop Treatment by pressing the Start/Stop 
(Emergency Stop) button.
Comfort Zone™
If the patient experiences discomfort, the 
Comfort Zone button may be pressed 
and the carriage will immediately reverse 
direction and move toward the extension 
angle. For the next five cycles, the carriage 
will operate in a range 5° less than the angle 
where the button was pressed. Once five 
cycles are complete, the Comfort Zone 
Flexion Angle will increase by 1° each cycle 
until it reaches the Programmed Flexion 
Angle.
Emergency Stop
Should the patient need to stop the 
treatment , they may press the Start/Stop 
(Emergency Stop) button once and 
treatment will immediately stop.