O P E R A T I N G I N S T R U C T I O N S

BEFORE USING THE LIFECHOICE POC FOR THE FIRST TIME

The LifeChoice POC and external battery
should be charged for a minimum of 6 hours
before initial use. Connect the external battery
to the LifeChoice POC, connect the AC
adapter to the external battery and plug the
AC adapter into an electrical outlet.

Make sure to verify the green LED light above
the Battery Check is solid green. This indicates
the system recognizes the external power source.
If the external battery is plugged into the Life-
Choice POC and the green LED light above the
Battery Check does not turn solid, the source is
not recognized. If charging the unit through the
external battery when this occurs, the charge will
not reach the internal battery. This may be an
indication the external battery needs to be reset.

Instructions for resetting the battery are provided in the Powering the LifeChoice POC section.

When more than two hours battery operation is needed, the LifeChoice POC should be
powered using the external battery first and then the internal battery. The LifeChoice POC
will automatically switch to the internal battery once the external battery has been depleted.

The LifeChoice POC is recommended and
designed to be operated and stored in its carry
case at all times.

Verify that the intake vents are clean and free
from obstruction.

Connect the nasal cannula securely to the
LifeChoice POC cannula nozzle fitting.