Important, Ix71/ix51, Precautions when unpacking the microscope – Olympus IX71 User Manual

Page 5

Advertising
background image

1

IX71/IX51

IMPORTANT

This microscope employs a UIS2/UIS (Universal Infinity System) optical design, and should be used with
modules designed for the IX2 or BX2 series (as well as some of the modules designed for the Olympus
IX or BX series).
For the applicable modules, please consult Olympus or the latest catalogues. Less than optimum per-
formance may result if inappropriate module combinations are used.
If you want to use one or more motorized modules, you should prepare the IX2-UCB control box, U-
HSTR2 hand switch, a PC and the IX2-BSW (Ver. 01.03) software (which runs on Windows 2000 or Win-
dows Me).

Configuration of Instruction Manuals

Since these microscopes are expandable to a variety of systems, separate instruction manuals are prepared
so that the user has to read only the manuals according to the user’s own system.

Manual Name

Main Contents

IX71/IX51

Observation procedures including transmitted light brightfield,
phase contrast and DIC observations

Fluorescence System for IX2

Reflected light fluorescence observation

IX2-GCP

Glass Center Plate

IX2-MLWCD

Mid-Long Working Distance Condenser

IX2-DICD

DIC Condenser

U-FWT/FWR/FWO

Motorized filter wheels (The U-FWT cannot be used with these
microscopes.)

IX2-UCB/U-HSTR2

Functions of the Control Box (incorporating the power supply)
and Hand Switch

IX2 Software for PC (CD-ROM)
IX2-BSW (Ver. 01.03 or later)

Methods of PC control of microscope functions

Precautions When Unpacking the Microscope

Releasing the Transport Lock of the Revolving Nosepiece

# Never attempt to rotate the coarse or fine adjustment knob @ without

removing the clamping rod. Otherwise, the focusing mechanism may
be damaged.

1. Loosen the screw ³ of the clamping rod ² using the Allen screwdriver

provided with the microscope frame.

2. Rotate the coarse and fine adjustment knobs @ in the direction of the

arrow and remove the clamping rod ².

3. Attach the provided seal (10 mm dia., black) on the hole made after remov-

ing the transport lock knob to prevent penetration of dust through the hole.

}Retain the clamping rod ² and screw ³ carefully because they will be

used again the next time the microscope is transported.

Releasing the Transport Lock of the Light Path Selector

# Never attempt to operate the light path selector without removing the

transport lock knob |. Otherwise, the light path selector mechanism
may be damaged.

· Rotate the knob counterclockwise to remove it.
· To prevent penetration of dust through the hole made after removing the

transport lock knob, stop the hole by attaching the provided seal (10 mm
dia., black).

}Retain the knob carefully because it will be used again the next time the

microscope is transported.

Stage (IX2-SFR, IX-MVR)

· Before transporting the stage, fix the flexible knobs with pieces of

adhesive tape so that they will not move.

Fig. 1

@

²

³

|

Advertising
This manual is related to the following products: