3 alarm configuration commands and queries, Alarm configuration commands and queries, Remote interface reference – American Magnetics 187 Self-Compensating Liquid Level Controller User Manual

Page 54

Advertising
background image

43

Remote Interface Reference

Alarm Configuration Commands and Queries

sum of the bits you wish to enable. Refer to Table 4-3 on page 36 for more
information. For example, to enable all alarms to be reported in bit 2 of the
Status Byte register, send:

STATus:ALARm:ENABle 127;

W

STATus:ALARm:ENABle?

Returns a decimal sum which corresponds to the binary-weighted sum of
the bits enabled by the last

STATus:ALARm:ENABle

<

enable_value

>

command.

4.5.3

Alarm Configuration Commands and Queries

The alarm configuration commands and queries provide read/write access
to the alarm settings available within the HI, A, B, LO, RATE, and MUTE
menus.

W

CONFigure:ALARm:HI <

level

>

Sets the HI alarm limit in the presently selected remote units.

W

ALARm:HI?

Returns the HI alarm value in the presently selected remote units.

W

CONFigure:ALARm:A <

level

>

Sets the A alarm setpoint in the presently selected remote units.

W

ALARm:A?

Returns the A alarm setpoint in the presently selected remote units.

W

CONFigure:ALARm:B <

level

>

Sets the B alarm setpoint in the presently selected remote units.

W

ALARm:B?

Returns the B alarm setpoint in the presently selected remote units.

W

CONFigure:ALARm:LO <

level

>

Sets the LO alarm limit in the presently selected remote units.

W

ALARm:LO?

Returns the LO alarm value in the presently selected remote units.

W

CONFigure:ALARm:RATE <

rate

>

Sets the RATE alarm limit in the presently selected remote units per
minute. The RATE alarm applies to both positive (fill) or negative (empty)
usage rates.