Installation – American Magnetics 188CPS Point-Sensing Instrument User Manual
Page 13
 
12
Installation
Installing the sensor
2.
Rack mount the instrument if desired
If the instrument has a rack mount chassis, follow the following procedure:
a.
Attach the rack mount adapter pieces to the instrument by first removing 
the four screws on the side of the instrument that attach the cover to the 
chassis. Attach the rack mount adapter pieces to the sides of the 
instrument by reinstalling the screws.
b.
Install the monitor in a 19" rack by securing the front panel to the rail in 
each of the four corners with mounting hardware supplied by the cabinet 
manufacturer.
3.
Install the sensor in the target vessel
Exercise care when installing the sensor since dents, crimps, bends or other physical 
distortions in the plate sensors will change electrical characteristics possibly causing 
calibration errors and/or disruption of proper instrument operation. Before installing 
the sensor, the user may want to review the Calibration and Operation sections to 
determine what, if any, calibration procedures may be necessary.
Note
The coaxial interconnecting cables and the oscillators are temperature 
sensitive and should be mounted in such a manner as to avoid very large 
temperature variations such as those encountered in the path of dewar vents. 
Large temperature gradients (for tank penetrations) are acceptable as long as 
the gradient is relatively constant during operation.
Make a note of the physical location of each point-sensing plate pair on the sensor 
assembly, and the physical location of the entire sensor assembly with respect to the 
vessel. This information will be necessary to associate an LED indication from the 
instrument with an actual liquid level.
4.
Connect the oscillator cables to the AMI sensor
Connect the oscillators to the sensor using the supplied RG-59/U coaxial cable. 
Ensure the oscillator is connected in the correct orientation (see page 5 for a system 
diagram). The cable length between the oscillator and the sensor should not exceed 6 
feet if possible.
Mark each oscillator cable with a labeling scheme that identifies a cable with a 
specific plate pair and its associated location on the sensor assembly.
Some sensor assemblies may integrate a continuous sensing element with several plate 
sensors. Do not connect the continuous sensing element to a Model 188CPS 
instrument. The Model 188CPS is designed only for point-sensing application.