Installation

Power-Up Procedure

magnet, AMI has preset the setup menu values for proper

operation. See sections 3.3, 3.5, 3.9 and 3.10 for more discussion of

the setup menu values and their entry into the Model 430

Programmer.

7. Set the Model 430 Programmer to display current (rather than

field). Refer to sections 3.2.1 and 3.7.5.

8. Set the ramp rate to 1 A/sec. Refer to sections 3.3 on page 25 and

3.7.1 on page 33.

9. Set the target current to 10 A. Refer to sections 3.3 on page 25 and

3.6.2 on page 31.

10. If a Persistent Switch is installed, set the PSw P/S Ramp Rate to 10

A/sec. Refer to paragraph 3.10.2.11 on page 56

11. Initiate ramping to the target current by pressing the

RAMP /

PAUSE

key (status indicator changes from

P

to ).

12. The system should ramp to 10 A in approximately 10 seconds.

Verify this is the case (if a PSwitch is installed and in the cooled

state, ramp time to 10 A should be slightly less than 2 seconds).

13. When the target current is achieved, the

FIELD AT TARGET

LED

will be illuminated. The display should show “

+10.00 A -

indicating that the Model 430 Programmer is in the holding mode

at the target current value (+10.00 A).

Note

There may be a discrepancy between the current shown on the power

supply display

1

and the current displayed on the Model 430

Programmer. The current measurement system incorporated in the

Model 430 is more accurate than the power supply shunt.

14. Verify that the power supply output current display indicates that

a total of approximately 10 A is being supplied to the load (which is

only the cabling in this case).

15. Set the target current to the Current Limit value. Refer to section

3.10.2.4 on page 52 to determine the Current Limit value. After the

new target current value is entered, the Model 430 Programmer

should ramp automatically to the new setting.

16. When the new target current value is reached, the power supply

current display (if provided) should also indicate the new value.

1. Not all power supplies have a local current readout.