14
Gas sampling and gas conditioning technology
6-1.1.10-ME
14.1
CHANGING THE DIAPHRAGM, VALVES AND SEALINGS
Hexagon screws
Figure 3
Pump head with hexagon screws
Loosen the 4 hexagon screws;
Remove the 2 parts of the pump head.
Valve plates
Sealing rings
Diaphragm
Figure 4
Dismounted pump head
Unscrew the diaphragm by hand out of the tapping hole of the rod (counter-clockwise);
Screw the new diaphragm into the rod hand-tight;
Change the sealing rings and valve plates;
Put on the basic part of the pump head and control that the bulge of the diaphragm fits to the
groove of the pump head;
Put on the upper part of the pump head so that its pin fits into the hole in the basic part of the pump
head;
Fix the 4 hexagon screws constantly over cross;