Maintenance – M&C TechGroup FM-200K-H_FA Operator's manual User Manual
Page 13

13
Gas sampling and gas conditioning technology
5-6.5.1-ME
18
MAINTENANCE
N O T E !
The safety instructions specific to the plant and process are to be
consulted prior to any maintenance work!
It is difficult to give any recommendations as to a particular maintenance cycle. Depending on your
process conditions, a meaningful maintenance cycle must be elaborated for the specific application.
Check from time to time, if the flowmeter glass is soiled.
W A R N I N G !
Aggressive condensate is .
Wear protective glasses and proper protective clothing!
High surface temperatures! Wear protective gloves!
To clean the flowmeter glass remove cover lid.
Unscrew the aluminium lid.
Disconnect the thermal conductivity jaws on the top and at the bottom of the flowmeter in two
parts by loosening the screws, two each.
Disconnect the 2 sample lines.
Disconnect the tube at the top of the flowmeter.
Take out the flowmeter including needle valve.
Loosen the 3 hexagon screws on the upper and lower adapter, until you can pull them out.
Pull out the glass, check and clean it.
Check O-rings on top and bottom and change if necessary.
Assemble the flowmeter in reverse sequence.
N O T E !
Prior to carrying out maintenance work on electrical parts, mains
voltage should be disconnected from all poles!
This also applies to any external control circuits, which may be
connected.