Differential pressure and t, Time, Combinations of options – M&C TechGroup SP3110 Data sheet User Manual
Page 4

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.11
∆P and T90 at flow of:
100
200
500
1000
1500
Nl/hr
∆P with new filter element F-3SS150
0,006
0,012
0,040
0,110
0,215
bar
∆P with new filter element S-2K150
0,003
0,005
0,02
0,058
0,135
bar
T90-time for SP3110 without tube
6
3,5
1
<0,5
<0,5
s
Dimensions
SP3110(V) basic version with option back purging type RS and heating type HEX4
Dimensions in mm
540
200
133
360
440
270
407
340
140
40
207
RS
HEX4
Differential pressure and T
90
-time
X = possible combinations
X
= recommended combinations
Combinations of options
Technical specifications and illustrations are without
obligation, subject to modifications. 10.04/11.12
Option No.
1
2
3
4
5
6
7a
7b
8
9
10
/RS
/VA
/3VA
/MS1
/R
/R/CC(/CCF) /BB
/BBF
/C/I
/2X
/FW
/RS
X
X
X
/VA
X
X
X
X
X
X
X
X
/3VA
X
X
X
/MS1
X
X
/R
X
/R/CC(CCF)
X
X
X
/BB
X
X
X
X
/BBF
X
X
X
/C/I
X
X
X
X
X
X
/2X
X
X
X
X
X
X