Figure 4, Mounting of the sensor la 1s into flow chamber ls – M&C TechGroup LA 1.4 Operator's manual User Manual

Page 13

Advertising
background image

13

Gas sampling and conditioning technology

5-5.1.1-ME

Union nut

Antikink

Item Part-No.

Description

1

90 E 1000

FPM O-ring 14

2

90 E 1010

PVDF

– pressing ring 16

3

03 E 1001

Liquid sensor LA 1S

with 4m connection cable

4

03 E 3100

Flow chamber LS

with union nut

Figure 4

Mounting of the sensor LA 1S into flow chamber LS


12.4 MOUNTING OF THE SENSOR LA 25S ONTO THE UNIVERSAL FILTER

If the client hat selected single components, the mounting of the sensor has to be effected as follows:

Screw the attached mounting fitting with a Teflon tape into the filter body.

Put the sensor with pre-assembled union nut and the pre-assembled pressing and cutting ring
into the fitting and draw up the union nut hand-tight. Then draw the union nut with a flat key
exactly 1 ¼ rotations. Now, it is mounted correctly.


12.5 MOUNTAGE OF THE SENSOR LA 25S INTO THE FLOW CHAMBER LS 25

If the client has selected single components, the mounting of the sensor has to be effected as follows:

Put the sensor with pre-assembled union nut and the pre-assembled pressing and cutting ring
into the flow chamber and draw the union nut hand-tight. Then draw the union nut with a flat
key exactly 1 ¼ rotations. Now, it is mounted correctly.