Figure 4, Mounting of the sensor la 1s into flow chamber ls – M&C TechGroup LA 1.4 Operator's manual User Manual
Page 13

13
Gas sampling and conditioning technology
5-5.1.1-ME
Union nut
Antikink
Item Part-No.
Description
1
90 E 1000
FPM O-ring 14
2
90 E 1010
PVDF
– pressing ring 16
3
03 E 1001
Liquid sensor LA 1S
with 4m connection cable
4
03 E 3100
Flow chamber LS
with union nut
Figure 4
Mounting of the sensor LA 1S into flow chamber LS
12.4 MOUNTING OF THE SENSOR LA 25S ONTO THE UNIVERSAL FILTER
If the client hat selected single components, the mounting of the sensor has to be effected as follows:
Screw the attached mounting fitting with a Teflon tape into the filter body.
Put the sensor with pre-assembled union nut and the pre-assembled pressing and cutting ring
into the fitting and draw up the union nut hand-tight. Then draw the union nut with a flat key
exactly 1 ¼ rotations. Now, it is mounted correctly.
12.5 MOUNTAGE OF THE SENSOR LA 25S INTO THE FLOW CHAMBER LS 25
If the client has selected single components, the mounting of the sensor has to be effected as follows:
Put the sensor with pre-assembled union nut and the pre-assembled pressing and cutting ring
into the flow chamber and draw the union nut hand-tight. Then draw the union nut with a flat
key exactly 1 ¼ rotations. Now, it is mounted correctly.