Closing down – M&C TechGroup PMA 100-L Operator's manual User Manual

Page 27

Advertising
background image

9-3.12.1-ME

M&C Products Analysentechnik GmbH

- 27 -

The following alarm messages are available at the common
alarm output:
• concentration alarm 1 - 4 (if allocated, see above)

Status messages like:
• flow alarm,
• transducer temperature,
• hardware fault, and
• leaving the current measuring range.

are signalised via the alarm LED, the analyser display and
the status contact output. Menu point 12.4.6 enables to
identify a status fault via a permanent mA-output signal. It
is available at the Sub-D socket X3 (see 8.2.1).
The mA-measuring signal is overlaid by the status signal
output.

16.Closing down

• In case of a temporary closing down of the process con–

trol systems around the analyser, the AC mains supply of
the analyser must remain „ON“.
No further precautions are required.

• In case of a closing down of the analyser for a longer

period, it is recommended to flush the analyser with clean
air in order to prevent that harmful gases will be left in
the instrument and damage the measuring cell.

Advertising