Use of the temperature controller, Display and function keys, Operating state of the temperature controller – M&C TechGroup BA-P4 Operator's manual User Manual
Page 22

22
Gas sampling and conditioning technology
10-1.1-ME
Heating up
Heating flashes
Ramp function
Sensor control
16
USE OF THE TEMPERATURE CONTROLLER
16.1
DISPLAY AND FUNCTION KEYS
Process value display, red
3
4 9 9 . 5
500.00
2
1
PGM
EXIT
Set-point display, green
PGM key
- to select the parameters
Decrement key
- to alter values
Increment key
- to alter values
EXIT key
- to quit the levels
LED for ramp/programme function, green
LED for status indication, yellow
- outputs 1 to 3
16.2
OPERATING STATE OF THE TEMPERATURE CONTROLLER
The operating state is displayed via the LED‟s 1 to 3 like:
Operating state
1
2
3
Heating up
off
on
blinking
Normal
heating on
on
on
on
heating off
on
on
off
Alarm
X
off
X
In case of a high or low temperature alarm, the controller passes automatically into the state of self-
locking heater switch off.
Setting value reached
Operating
temperature
or setting value reached