Technical data, 8technical data – M&C TechGroup PMA 20 Operator's manual User Manual

Page 8

Advertising
background image

8

Gas sampling and gas conditioning technology

9-3.5.1-ME

8

TECHNICAL DATA

Oxygen analyser Series PMA

Version PMA 20

Part No.

02A1000 = 230V 50Hz, 02A1000a = 115V 60Hz

Measuring ranges

selectable for 0-3, 0-10, 0-30 and 0-100 vol.% O

2

, linear

Indication

analogue meter with a scale of 0-30 and 0-100% for each selected
range
analogue meter with a scale of 0-30 and 0-100% for each selected
range, digital meter, 3 1/2 digit 9 mm high
for 0-100%O

2

reading, selectivity 0,1vol.% O

2

Output signal

0-1V DC non-isolated, load > 100 K

Ω, for the range of 100 vol.%, and

0-20mA or 4-20mA* for the chosen range, non-isolated, max. load
300

Ω

Response time for 90% FSD

< 3 seconds at 60 Nl/hr air

Accuracy after calibration

deviation: analogue = ± 1% of span / digital = ± 0,1 vol.% O

2

Reproducibility deviation

analogue = < 1% of span / digital = ± 0,1 vol.% O2

Influence of ambient temperature

no influence up to 45 °C

Influence of barometric pressure

The oxygen reading varies in direct proportion to changes of the ba-
rometric pressure.

Influence of sample gas flow

variation in gas flow between 0-60 Nl/hr air will cause a difference of
< 0,1 vol.% O

2

.

Sample gas inlet pressure

0,01 up to 1 bar g, (PMA 20 required admission pressure for compe-
tent flow rate, no pump inside)

Sample gas outlet pressure

Outlet of analyser must discharge freely into atmosphere.

Flow rate of sample gas

max. 60 Nl/hr air, adjustable with needle valve on the flowmeter 7-70
Nl/hr

Temperature of sample gas

-10 °C up to +40 °C, dry gas

Analyser temperature

fixed at +50 °C

Ambient temperature

-10 °C up to +45 °C

Storage temperature

-20 °C up to +60 °C, relative humidity 0-90% RH

Power supply

internal power unit for 230VAC standard or 115VAC available (a)*
+/-10%, 40-60Hz, 26,5VA

Electrical connections

terminals 2,5 mm

2

; 1x PG11, 1x PG13,5 cable gland

Materials in contact with sample gas

Platinum, Glass, Polypropylene, Stainless Steel 316, FPM, Epoxy
resin

Sample gas connection

PP hose connectors DN 4/6 for 4 mm i.d. and 6 mm o.d. tube

Protection / electrical standard

IP 53 EN 60529 / EN 61010

Housing / colour

plastic wall mounting housing / blue/grey

Dimension / weight

height 220 mm, width 214 mm, length 182 mm / approx. 3 kg

Options

Extra charge for PMA 20 in solvent resistant version with special measuring cell type PMC-1LB.
Material: O-ring: Kalrez, flowmeter: PVDF, tubing and fittings: PVDF/SS316.
Extra charge for PMA 20 in chlorine resistant execution with a special measuring cell type PMC-
1CL2. All gas bearing tubes PTFE/PVDF equipped with purge gas connectors.
Extra charge for qualification test according to TA-Luft + EN 14181 resp. 13. and 17. BImSchV
Extra charge for PMA with combined analogue/digital display.
analogue meter with a scale of 0-30 and 0-100% for each selected range, digital meter, 3 1/2
digit 9 mm high
LCD-indicator for 0-100%O

2

reading, selectivity 0,1vol.% O

2

or digital meter 3 1/2 digit 18 mm

high LCD for 0-100% O

2

reading, selectivity 0,1 vol.% O

2







Advertising