Gas sampling and conditioning technology

9-0.2-ME

Figure 3

Notch to differentiate the two cartridge types

N O T E !


For catalyst type SS, oxygen in the sample gas as well as in the test gas (>1
Vol.% O

2

) is necessary. A calibration with NO

2

or N

2

is not possible.


8.2

SELECTION OF THE APPROPRIATE CATALYST TEMPERATURE

The converter with carbon-molybdenum cartridge filling is pre-adjusted to 350°C in the factory. In case
of a metallic cartridge filling, the converter temperature is adjusted to 660°C. The catalyst temperature
can be adjusted continuously using the temperature controller on the front of the converter (see 17.2).

N O T E !

Depending on the gas composition and the flow temperature can be
adjusted and optimised within a certain level:
320°C

– 380°C, type C respectively

620°C

– 680°C type SS.

N O T E !

High NO

2

-concentrations, especially in connection with high flow ra-

tios need higher operating temperatutres of the converter, in particu-
lar the start up process of the maximum values.

N O T E !

High operating temperatures of the converter, especially the start up
process of the maximum values leads to an increased abrasion of
converter components, in particular sealing- and oven-components!

D A N G E R !


The maximum allowable operating temperature for catalyst type C is
380°C. Temperatures above 400°C lead to a combustion of the conver-
ter material. Urgent fire danger !

D A N G E R !

The maximum allowable operating temperature for catalyst type SS is
680°C. Temperatures above 680°C can lead to a destruction of the
catalyst resp. converter!

Type SS