Specifications, Functions, Accessories – RBP Chemical Technology ULTRAPEN PT1 User Manual

Page 2: Myron l warranty

Advertising
background image

SPECIFICATIONS

Measurement Range:

1 - 9999 µS or ppm (0.0010 - 9.999 ppt salinity)

Accuracy (After Wet Calibration):

± 1% of reading

Repeatability:

< 1000 µS or ppm ± 1 Count
≥ 1000 µS or ppm ± 0.3% of reading

Resolution:

Conductivity and TDS:
0.1 for 1.0 - 99.9 µS or ppm
1 for 100 - 9999 µS or ppm

Salinity:
0.0001 for 0.0010 - 0.0999 ppt
0.001 for 0.100 - 9.999 ppt

Temperature:
0.1 ºC or ºF

Time to Reading Stabilization:

10 - 20 seconds

Active Mode Power Consumption:

30 - 100 mA

Sleep Mode Power Consumption:

2 µA

Temperature Measurement Range:

0 - 71 °C or 32 - 160 °F

Temperature Accuracy Displayed:

± 0.1 ºC or ± 0.2 ºF

Temperature Compensation Method:

Automatic to 25 ºC

Physical Dimensions:

17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D

Weight:

55 g or 1.94 oz.

Case Material:

Anodized Aluminum with UV Protective Coating

Battery Type:

N type, Alkaline

Battery Voltage:

1.5 V

Calibration Solution Point:

1800 µS KCl; 3000 ppm 442™ (2027 ppm NaCl)

Operating/Storage Temperature:

0 - 55ºC or 32 - 131ºF

Water Resistance:

IP67 and NEMA 6

EN61236-1: 2006 - Annex A: 2008

Electrostatic discharge to case of instrument may cause PT1 to spontaneously power
on. In this case, the PT1 will power off after several seconds.

FUNCTIONS

Mode

Parameter

Solution Model

Units

Cond

KCl

Conductivity

potassium chloride

microsiemens (µS)

tdS

442

Total Dissolved Solids (TDS)

442™ Myron L Natural Water Standard

parts per million (ppm)

tdS

NaCl

TDS

sodium chloride

ppm

SALt

442

Salinity

442™ Myron L Natural Water Standard

parts per thousand (ppt)

SALt

NaCl

Salinity

sodium chloride

ppt

ACCESSORIES

Standard Solutions for calibration:
MODEL#

S

: KCL-1800, 442-3000 (2027 ppm NaCl)

MYRON L WARRANTY

Excepting the battery, the Myron L PT1 Conductivity/TDS/
Salinity Pen is warrantied to be free from manufacture
defect for 1 year.

Built On Trust. Founded in 1957, the Myron L Company is one of
the world’s leading manufacturers of water quality instruments.
Because of our commitment to product improvement,
changes in design and specifications are possible. You have
our assurance any changes will be guided by our product
philosophy: accuracy, reliability, and simplicity.

2450 Impala Drive
Carlsbad, CA 92010-7226 USA
Tel: +1-760-438-2021
Fax: +1-800-869-7668 / +1-760-931-9189
www.myronl.com

Printed in U.S.A.

© Myron L Company 2011

PT1DS 06-11

Advertising