1 calibration check, 2 reference spectrum interface – Soft dB Opus Suite SLM 4-Ch Module User Manual

Page 54

Advertising
background image

51

8.7.1

Calibration Check

The calibration check is generally used in long term or permanent environmental system. It is

intended to periodically verify that a system is working properly and that the acquisition level is

adequate. To achieve this goal, an actuator has to be installed near the input sensor. Prior the

measurement, the actuator will be exited to provide a reference signal at the input and the obtained

reference spectrum will be saved. During the acquisition, the actuator will be activated periodically.

A comparison of the spectrum during the measurement and the reference spectrum, allows the user

to conclude on the proper operation of the acquisition system.

The generated signal should be loud enough to be way over the normal signal being measured.

Ideally, the generated signal should be a few dB below the maximum range of the input. For

example, the Larson Davis 426A12 actuator typically generates 91 to 97 dB at the input.

The following sections explain how to configure the generator and how to define the reference

spectrum.

8.7.2

Reference Spectrum Interface

Current spectrum on inputs

Averaging

Controls

Generator Setup

Advertising