Safety and warnings – Velleman PCS100A pcs100 - pcs500 - k8031 User Manual

Page 7

Advertising
background image

PCS500 / PCS100 / K8031

7

SAFETY and WARNINGS

Symbols displayed on the unit

!

Important safety information, see user manual.

The PC scopes are optically isolated from the PC, but even then it

is advisable to measure only safe devices.

Measurements should be avoided in case of polluted or very humid air.
One should also refrain from measuring conductors or installations that
use voltages that exceed 600Vrm s above ground level. CAT II indicates
conformity for measurements of domestic installations.

The maximum input voltage for the connections of the unit stands at 100Vp
(AC+DC)

Do NOT open the enclosure while performing measurements.

Remove all test leads before opening the enclosure in order to avoid
electrical shock.

Use a measuring probe with an insulated connector ( e.g. type
PROBE60S) when measuring voltages exceeding 30V.


Before making measurements and for safety reasons, it is important to know
some information about the measured unit.
Safe devices are:

Battery operated equipment

Equipment supplied via a transformer or adapter.


Unsafe devices are:

Equipment directly connected to mains (e.g. old TV sets)

Equipment that contains components that are d irectly connected to mains
(dimmers…).

It is advisable, when measuring on above equipment, to use a isolation
transformer.
Always be careful when measuring directly to the mains; please remember
that the ground of both channels is interconnected!

F

IMPORTANT: Prior to measuring high voltages, set the probe to the X10
position.

Advertising
This manual is related to the following products: