Veris Technologies OpticMapper - Operating Instructions User Manual

Page 57

Advertising
background image

Pub.# OM17-OpticMapper

6-23

Replacing microchip. If microchip fails, or if unit needs updated firmware, it will be necessary to
remove chip and replace with new one. Before chip removal and installation, touch grounded metal
to discharge your static electricity. Remove chip with pliers or chip puller. Install new chip, making
sure that notch in plastic housing is to the right. Be careful to not damage pins of new chip as it is
installed

Figure 9.4


Figure 9.5 a and b. Notch on chip must be to the right, matching notch on chip socket

3100 to 3150 Resistor Change. Because the coulter-electrode spacing on a 3100 model is
different than a 3150, there are unique constants applied to the EC measurement values for each
model. The EC Surveyor Instrument Signal Test Procedure #1 shows the different signal test load
readings for each. For 3100 (and 2000XA) models, a resistor is installed on the circuit board. For
3150 models the resistor is not used. When the EC Surveyor is purchased with a 2000XA or 3100
model, the resistor is installed, and removed for the 3150 models. If it becomes necessary to
change this in the field, disconnect power cord, remove cover as shown above and remove or
install resistor as shown below.

notch

notch

Advertising